4-34 Programming Site Data
Entering Analog Inputs
The Analog Inputs option enables you to specify parameters for
general purpose or standard volume inputs. To enter the Analog
Inputs option, scroll to the Analog Inputs entry on the Program
Menu and press
[ENTER]. If you have not enabled the Standard
Volume Option (see page 4-18), the screen appears similar to
Figure 4-15 below. To step through each parameter, press the [T]
key. Refer to Figure A-6 on page A-6 of Appendix A, Menu
Maps.
Entering General-Purpose Analog Inputs
Figure 4-15: The Analog Inputs Option Window
1. The first prompt asks you to select whether the desired
function is off or general purpose.
a. Press
[ENTER] to open the drop-down menu.
b. Scroll to the desired response.
c. Press
[ENTER] to confirm your selection.
Note: If you select “Off,” you cannot select any other prompts in
this window.
Analog Input
Cancell OK
Function
Units
Span
Label
Zero
100
General Purpose
0
Psia
Input A Input B
Page view 94