GE PT878GC User Manual Page 70

  • Download
  • Add to my manuals
  • Print
  • Page
    / 308
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 69
Programming Site Data 4-9
April 2004
Entering Pipe Parameters (cont.)
2. At this point, the program follows one of two paths:
a. If you have selected a preprogrammed material, the
PT878GC automatically supplies the correct sound speed,
and proceeds to step 3.
b. If you have selected “Other,” the meter prompts you to
enter the sound speed. Use the numeric keys to type the
known sound speed in the text box, and press
[ENTER] to
confirm the choice.
3. At this point, two alternatives are available. At the Diameter
prompt, the meter asks for the pipe outside diameter and
thickness. But if you have selected certain pipe materials
(carbon or stainless steel, cast iron, PVC and CPVC), you
have the option of entering the pipe dimensions by a
standardized schedule. Once you enter the nominal pipe size
and schedule identification, the PT878GC determines the OD
and wall thickness from an internal table.
If you select a material that uses the Diameter alternative:
a. You must select from two choices, outside diameter or
circumference.
The prompt moves to the OD (outside diameter) text box.
Type the pipe OD (in mm or in.) into the text box, and press
[ENTER] to confirm the choice, or
Move the prompt to the OD X PI (circumference) text box.
Type the circumference (in mm or in.) into the box, and
press
[ENTER] to confirm your choice.
b. In either case, the next prompt asks for the wall thickness.
Type the value (in mm or in.) into the box, and press
[ENTER] to confirm your choice
Note: The measurement units shown depend on the choices you
have made in the English/Metric window of the Meter
menu.
Page view 69