GE PT878GC User Manual Page 54

  • Download
  • Add to my manuals
  • Print
  • Page
    / 308
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 53
April 2004
3-22
Installing DMP-1 Dampening Material with CFG-V
Series Fixtures (cont.)
7. With a marker, trace around the transducer footprint, as shown
in Figure 3-24 below.
Figure 3-24: DMP-1 Material with
Transducer Footprint Traced
8. Remove the fixture and transducer. Then use a utility knife to
cut out the area under the transducer footprint, and peel the cut
material off the pipe, as shown in Figure 3-25 below.
Figure 3-25: DMP-1 Material with
Transducer Footprint Cut Out
9. To remove any residual adhesive, clean the cut-out area with a
rag or a piece of sandpaper.
10. Reinstall the fixture and transducers on the pipe.
Page view 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 307 308

Comments to this Manuals

No comments