4 - 20 mA Calibration
Advanced Setup
20 mA Calibration
To calibrate Span at 20 mA:
1. Click on 20mA (Figure 229).
Figure 229 Starting 20 mA Calibration
Once 20 mA calibration is started, the 20 mA button is greyed out.
Figure 230 20 mA Calibration Active
2. Read the value from the precision milliammeter.