ValVue Digital Communications Software Ver. 2.80
136
GE Oil & Gas
View Saved Diag
You can view previously saved diagnostic graphs and data. To view saved diagnostics:
1. Click View Saved Diag and an empty Diagnostic Graph appears.
2. Click Select Curve and a dialog containing a list of signature locations appears
(Figure 113).
Figure 113 Choose Signature Curve
3. Choose the type of saved signature curve for viewing, click OK and, if appropriate, a
Windows browser to locate the file appears.
4. Select a curve is selected, the graph appears in the Diagnostic Graph window.