Illumination System (Continued)
Borescope Body
The body is the hand-piece of the borescope. Focus, Zoom, Rotary Scan and Swing
Prism control knobs are conveniently located on the body.
Shaft (or Barrel)
The shaft, or barrel, is the working length of the borescope and carries the optical
train into equipment to be inspected. The shaft’s unique triple tube design provides
special protection of the optical train and enhances the durability of the scope.
Distal Section
The distal section houses the termination point of the illumination fiber-optic bundle and
the lens set that focuses images into the optical train. These components are assembled
in a precise orientation to provide the highest possible imaging performance, along with
the desired direction and field-of-view.
Setup consists only of focusing the eyepiece to your eye, plugging a light guide into the post
and available light source, and making a few checks to ensure the borescope is operating
properly.
Focus the Eyepiece: Remove protective lens caps and, if desired and available, attach the
rubber eyecup. Focus the eyepiece for your eye by rotating the focus knob while looking into
the eyepiece. The viewing tip should be pointed at an object with sharply defined edges. Typed
textonwhitepaperisarecommendedtargetforthisadjustment.
Plug into a Light Source:ConnecttheexitendofanACMIlightguidetotherigidscopepost
and insert the other end into the light guide socket of an Everest VIT or other compatible light
source. Turn the light source on, wait about 30 seconds for the lamp to reach peak intensity,
and check to ensure that a bright light is projected from the tip of the borescope. The scope
is now ready to use.
Check Operation: Look through the eyepiece to ensure the scope is providing a sharp, clear
image. Clean the eyepiece and distal cover glass if dirty. Check the operation of all equipped
features, such as focus, rotary scan, zoom and swing prism. If performance of any of these fea-
tures is degraded, return the scope to the Service Center for evaluation and repair. Continued
use of a damaged scope could cause further damage.