GE ISOVOLT Titan E Brochure Page 6

  • Download
  • Add to my manuals
  • Print
  • Page
    / 7
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 5
Technical Specifications
Unipolar Systems
High Voltage Generator
Max. Output Voltage 160 kV 225 kV
Max. Output Current 45 mA 45 mA
Max. Output Power 4,5 kW, Limited by Tube Specification 4,5 kW, Limited by Tube Specification
High Voltage Ripple 5 V/mA (With High Voltage Cable 10 m), 40 kHz 5 V/mA (With High Voltage Cable 10 m), 40 kHz
Insulation Oil Oil
Housing Dimensions (Cathode) (W x D x H) 350 x 870 x 850 mm (13.8” x 34.3” x 33.5”) 350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
Weight (Cathode) 189 kg (417 lbs), Including Power Module 189 kg (417 lbs), Including Power Module
Tube Voltage
Preselection and Setting From 5 to 160 kV in 0.1 kV / 1 kV / 10 kV From 5 to 225 kV in 0.1 kV / 1 kV / 10 kV
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 kV 0.1 kV
Accuracy < ±1% < ±1%
Reproducibility ±0.01% at Constant Temperature Level ±0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Tube Current
Preselection and Setting From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA From 0.1 to 45 mA in in 0.01 mA / 0.1 mA / 1 mA
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 mA / 0.01 mA 0.1 mA / 0.01 mA
Accuracy ± 1% ±1%
Reproducibility ± 0.01% at Constant Temperature Level ± 0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Exposure Time
Programmable Timer Non-Volatile Memory Non-Volatile Memory
Preselection and Setting From 0.1 to 99.9 Minutes in 0.1 Min. Increments or
from 1 to 999 Sec. in 1 Sec. Increments or as direct
Min./Sec. value (up to 99’59”)
From 0.1 to 99.9 Minutes in 0.1 Min. Increments or
from 1 to 999 Sec. in 1 Sec. Increments or as direct
Min./Sec. value (up to 99’59”)
Digital Display of Set and Actual Values The Remaining Time Is Displayed, i.e. After a Mains
Failure Exposure Can Be Continued Without any
Time Error
The Remaining Time Is Displayed, i.e. After a Mains
Failure Exposure Can Be Continued Without any
Time Error
Prewarning Time
Preselection and Setting Digital Setting From 2 to 250 Seconds or
de-activated
Digital Setting From 2 to 250 Seconds or
de-activated
Programmed Mode
Number of Storable Programs 250 250
Warm-Up Automatic Intelligent Tube conditioning based on
Real-time Clock
Extended warm-up for special conditioning
Automatic Intelligent Tube conditioning based on
Real-time Clock
Extended warm-up for special conditioning
X-ray Tube Setup 8 Tube selectable from a database of > 40
pre-programmed tubes
8 Tube selectable from a database of > 40
pre-programmed tubes
Operation History 512 Events (256 On/Off Events) 512 Events (256 On/Off Events)
Warm-up History 128 Events 128 Events
Control Module
Dimensions (W x D x H) 460 x 270 x 100 mm (18.1” x 10.6” x 3.9”)
Built into Desk Housing
460 x 270 x 100 mm (18.1” x 10.6” x 3.9”) Built into
Desk Housing
Weight 4.9 kg (10.8 lbs) Including Desk Housing 4.9 kg (10.8 lbs) Including Desk Housing
Connected Loads
Power connection 1N PE 230 V ± 10% 50/60 Hz
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase,
Grounded Neutral TN-S or TN-C-S Mains (Star
Connected System, Optional 3-Phase Isolation
Transformer)
1N PE 230 V ± 10% 50/60 Hz
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase,
Grounded Neutral TN-S or TN-C-S Mains (Star
Connected System, Optional 3-Phase Isolation
Transformer)
Grounding Separate Grounding for X-ray Tube and High
Voltage Generator (Minimum 6 mm
2
)
Separate Grounding for X-ray Tube and High
Voltage Generator (Minimum 6 mm
2
)
Mains Fuses 63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses,
Customer-Supplied
63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses,
Customer-Supplied
Operating Temperature Range C to +40°C C to +40°C
Storage Temperature Range -30°C to +70°C -30°C to +70°C
Bipolar Systems
High Voltage Generator
Max. Output Voltage -160 kV (Cathode), +160 kV (Anode) -225 kV (Cathode), +225 kV (Anode)
Max. Output Current 45 mA 45 mA
Max. Output Power 4,5 kW (Cathode)
3 kW (Anode)
Limited by Tube Specification
4,5 kW (Cathode)
3 kW (Anode)
Limited by Tube Specification
High Voltage Ripple 10 V/mA (With High Voltage Cable 10 m), 40 kHz 10 V/mA (With High Voltage Cable 10 m), 40 kHz
Insulation Oil Oil
Operation History 512 Events (256 On / Off events) 512 Events (256 On / Off events)
Warm-up History 128 Events 128 Events
Housing Dimensions (Cathode) (W x D x H) 350 x 870 x 620 mm (13.8” x 34.3” x 24.4”) 350 x 870 x 620 mm (13.8” x 34.3” x 24.4”)
350 x 870 x 850 mm (13.8” x 34.3” x 33.5”) 350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
Weight (Anode) 123 kg (272 lbs) 123 kg (272 lbs)
Weight (Cathode) 189 kg (417 lbs), Including Power Module 189 kg (417 lbs), Including Power Module
Tube Voltage (Anode)
Preselection and Setting From 5 to 320 kV in 0.1 kV / 1 kV / 10 kV From 5 to 450 kV in 0.1 kV / 1 kV / 10 kV
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 kV 0.1 kV
Accuracy < ±1% < ±1%
Reproducibility ±0.01% at Constant Temperature Level ±0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Tube Current
Preselection and Setting From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA
Digital Display of Set and Actual Values Simultaneous 4 Digits Each Simultaneous 4 Digits Each
Display Resolution 0.1 mA / 0.01 mA 0.1 mA / 0.01 mA
Accuracy ± 1% ±1%
Reproducibility ± 0.01% at Constant Temperature Level ± 0.01% at Constant Temperature Level
Temperature Drift < 65 ppm/°C < 65 ppm/°C
Exposure Time
Programmable Timer Non-Volatile Memory Non-Volatile Memory
Preselection and Setting From 0.1 to 99.9 Minutes in 0.1 Min. Increments or
From 1 to 999 Sec. in 1 Sec. Increments or as
direct Min./Sec. value (up to 99’59”)
From 0.1 to 99.9 Minutes in 0.1 Min. Increments or
From 1 to 999 Sec. in 1 Sec. Increments or as
direct Min./Sec. value (up to 99’59”)
Digital Display of Set and Actual Values The Remaining Time Is Displayed, i.e. After a Mains
Failure Exposure Can Be Continued Without any
Time Error
The Remaining Time Is Displayed, i.e. After a Mains
Failure Exposure Can Be Continued Without any
Time Error
Prewarning Time
Preselection and Setting Digital Setting From 2 to 250 Seconds or
de-activated
Digital Setting From 2 to 250 Seconds or
de-activated
Programmed Mode
Number of Storable Programs 250 250
Warm-Up Automatic Intelligent Tube conditioning based on
Real-time Clock
Extended warm-up for special conditioning
Automatic Intelligent Tube conditioning based on
Real-time Clock
Extended warm-up for special conditioning
X-ray Tube Setup 8 Tube selectable from a database of > 40
pre-programmed tubes
8 Tube selectable from a database of > 40
pre-programmed tubes
Control Module
Dimensions (W x D x H) 460 x 270 x 100 mm (18.1” x 10.6” x 3.9”)
Built into Desk Housing
460 x 270 x 100 mm (18.1” x 10.6” x 3.9”)
Built into Desk Housing
Weight 4.9 kg (10.8 lbs) Including Desk Housing 4.9 kg (10.8 lbs) Including Desk Housing
Connected Loads
Power connection 1N PE 230 V ± 10% 50/60 Hz
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase,
Grounded Neutral TN-S or TN-C-S Mains (Star
Connected System, Optional 3-Phase Isolation
Transformer)
1N PE 230 V ± 10% 50/60 Hz
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase,
Grounded Neutral TN-S or TN-C-S Mains (Star
Connected System, Optional 3-Phase Isolation
Transformer)
Grounding Separate Grounding for X-ray Tube and High
Voltage Generator (Minimum 6 mm
2
)
Separate Grounding for X-ray Tube and High
Voltage Generator (Minimum 6 mm
2
)
Mains Fuses 63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses,
Customer-Supplied
63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses,
Customer-Supplied
Operating Temperature Range 0°C to +40°C C to +40°C
Storage Temperature Range -30°C to +70°C -30°C to +70°C
1110
Page view 5
1 2 3 4 5 6 7

Comments to this Manuals

No comments